苏明旭, Muhammad Abdul Ahad, 蒋永, 等. 超声脉冲反射法和激光吸收光谱法同步测量流动液膜厚度[J]. 化工学报, 2018,(7):2972-2978.
SU Mingxu, MUHAMMAD Abdul Ahad, JIANG Yong, et al. Synchronous thickness measurements of flowing liquid film on horizontal surface by ultrasonic pulse-echo and laser absorption spectroscopy methods[J]. CIESC Journal, 2018, (7): 2972-2978.
SU Mingxu, MUHAMMAD Abdul Ahad, JIANG Yong, et al. Synchronous thickness measurements of flowing liquid film on horizontal surface by ultrasonic pulse-echo and laser absorption spectroscopy methods[J]. CIESC Journal, 2018, (7): 2972-2978.DOI: 10.11949/j.issn.0438-1157.20171251.
Because of common existence of liquid film flow in industrial processes
it's crucial to study methods for measuring thickness of moving liquid film. Measurement accuracy of both ultrasonic pulse-echo and laser absorption spectroscopy methods was validated by calibration standards of known film thicknesses (100-1000 μm). The coefficient of variation was 1.07% for ultrasonic pulse-echo method and 1.29% for laser absorption spectroscopy method. Then
thickness of flowing liquid film was measured by these two methods. The average film thicknesses were in good agreement for two methods when liquid film flow was at low/medium/high velocity. At three corresponding velocities
the difference of average film thicknesses between these two methods were 16.59
16.26 and 13.36 μm
and the relative standard deviation of film thickness was 0.29%
7.71% and 25.37%
respectively. Furthermore
both methods detected same number of fluctuations of liquid film within 1 s cycle at three different velocities.